《分流不分流進樣口介紹》由會員分享,可在線閱讀,更多相關(guān)《分流不分流進樣口介紹(14頁珍藏版)》請在裝配圖網(wǎng)上搜索。
1、Slide #16850/6890 Split AnimationSplit/Splitless Inlet - Component IdentificationSeptumLinerLiner O-RingGold SealCapillary Column5/4/00Slide #26850/90 Split/Splitless Inlet - Component IdentificationSeptumSEPTUM PURGEVENTInlet SplitPortSlide #36850/90 Split Animation - Default ConditionsFLOWSENSORPR
2、OPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORPURGEVALVEOPENTOTAL FLOWCONTROL LOOPSPLITVENTGOLD SEALINPUTPRESSUREPROPORTIONALVALVE 280 PSI0 ML/MINSPLIT VENTTRAP0PSISlide #46850/90 Split Animation - Total Flow SettingFLOWSENSORPROPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORPURGEVALVEOPEN
3、TOTAL FLOWCONTROL LOOPSPLITVENT205 ML/MINGOLD SEALINPUTPRESSUREPROPORTIONALVALVE 280 PSIPRIMARY FLOW PATHREDUCED FLOW PATH205ML/MINSPLIT VENTTRAP0PSISlide #56850/90 Split Animation - Inlet Pressure SettingFLOWSENSORPROPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTA
4、L FLOWCONTROL LOOPGOLD SEALINPUTPRESSUREPROPORTIONALVALVE 280 PSIPRIMARY FLOW PATHREDUCED FLOW PATH205ML/MINSPLIT VENTTRAP10PSISPLITVENT200 ML/MINCOLUMN FLOW2 ML/MINPRESSURIZED GASBACK PRESSURE CONTROL LOOPSlide #66850/90 Split Ready for Injection AnimationFLOWSENSORPROPORTIONALVALVE 1PRESSURESENSOR
5、SEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPINPUTPRESSUREPROPORTIONALVALVE 280 PSI205ML/MINSPLIT VENTTRAP10PSISPLITVENT200 ML/MINCOLUMN FLOW2 ML/MINBACK PRESSURE CONTROL LOOPSplit Ratio - 100:1Mode:SplitTemp 250250Pressure 10.010.0Split Ratio 100 Split Flow 200Tot Flow 205.
6、0205.0Gas SaverOFFSlide #76850/90 Split Insert Syringe AnimationFLOWSENSORPROPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPGOLD SEALINPUTPRESSUREPROPORTIONALVALVE 280 PSI205ML/MINSPLIT VENTTRAP10PSISPLITVENT200 ML/MINCOLUMN FLOW2 ML/MINBACK PRESS
7、URE CONTROL LOOPSplit Ratio - 100:1Slide #86850/90 Split Inject Sample AnimationFLOWSENSORPROPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPGOLD SEALINPUTPRESSUREPROPORTIONALVALVE 280 PSI205ML/MINSPLIT VENTTRAP10PSISPLITVENT200 ML/MINCOLUMN FLOW2
8、ML/MINBACK PRESSURE CONTROL LOOPSplit Ratio - 100:1Slide #96850/90 Split Sample Vaporizes AnimationFLOWSENSORPROPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPGOLD SEALINPUTPRESSUREPROPORTIONALVALVE 280 PSI205ML/MINSPLIT VENTTRAP10PSISPLITVENT200
9、ML/MINCOLUMN FLOW2 ML/MINBACK PRESSURE CONTROL LOOPSplit Ratio - 100:1Glass WoolSlide #106850/90 Split Sample is Split - Animation Good Liner Design?FLOWSENSORPROPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPINPUTPRESSUREPROPORTIONALVALVE 280 PSI
10、205ML/MINSPLIT VENTTRAP10PSISPLITVENT200 ML/MINCOLUMN FLOW2 ML/MINBACK PRESSURE CONTROL LOOPSplit Ratio - 100:1Smaller O.D. LinerOpen Split Flow PathGlass PositioningBeadGlassWoolTapered LinerAgilent SplitLiner - 5183-4647Slide #116850/90 Split Split Continues AnimationFLOWSENSORPROPORTIONALVALVE 1P
11、RESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPINPUTPRESSUREPROPORTIONALVALVE 280 PSI205ML/MINSPLIT VENTTRAP10PSISPLITVENT200 ML/MINCOLUMN FLOW2 ML/MINBACK PRESSURE CONTROL LOOPSplit Ratio - 100:1Slide #126850/90 Split Split is Complete AnimationFLOWSENSORPROPORTI
12、ONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPINPUTPRESSUREPROPORTIONALVALVE 280 PSI205ML/MINSPLIT VENTTRAP10PSISPLITVENT200 ML/MINCOLUMN FLOW2 ML/MINBACK PRESSURE CONTROL LOOPSplit Ratio - 100:1Slide #136850/90 Split Sample is separated Animation and
13、 DetectedFLOWSENSORPROPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPINPUTPRESSUREPROPORTIONALVALVE 280 PSI205ML/MINSPLIT VENTTRAP10PSISPLITVENT200 ML/MINCOLUMN FLOW2 ML/MINBACK PRESSURE CONTROL LOOPSplit Ratio - 100:1Slide #146850/90 Split Gas Sa
14、ver - Set to 55 ml/min Animation After the Injection is completeFLOWSENSORPROPORTIONALVALVE 1PRESSURESENSORSEPTUM PURGEREGULATORFIXED 3ML/MINPURGEVALVEOPENTOTAL FLOWCONTROL LOOPINPUTPRESSUREPROPORTIONALVALVE 280 PSI55ML/MINSPLIT VENTTRAP10PSISPLITVENT50 ML/MINCOLUMN FLOW2 ML/MINBACK PRESSURE CONTROL LOOPSplit Ratio - 100:1Mode:SplitTemp 250250Pressure 10.010.0Split Ratio 100 Split Flow 200Tot Flow 205.0205.0Gas SaverONSaver Flow50.0Saver Time0.5